Our current single-particle microbeam
system features an extensively upgraded spatial resolution of 0.6 μm
(diameter) for the beam. Complementing this, point and shoot targeting
system allows us to move the beam instead of the sample for faster
irradiations.
Our
multiphoton microscope is online to image 3D
tissue samples and small organisms.
Installation of our
laser ion source provides more
densely-ionizing radiations, i.e., heavier ion beams, with stopping powers
from 10 to 4,500 keV/μm, compared with
our current beams of H and He ions, which have stopping powers from 10 to
200 keV/μm. A
secondary electron ion
microscope is available to assess beam quality.
We have added x rays to the list of
microbeams we can produce.
Point and Shoot
 |
|
|
Photograph of
the Point and Shoot deflector coil. |
|
During microbeam irradiation, cells to
be irradiated are moved to the beam position using a high-resolution
three-axis piezo-electric stage. When a collimated microbeam was being
used, this was a necessary but relatively time-consuming method to
position cells for irradiation. Unlike a collimated microbeam, a focused
microbeam is not restricted to a single location on the exit window and
therefore can be deflected to any position in the field of view of the
microscope used to observe the cells during irradiation. Moving the beam
to the cell position magnetically or electrostatically can be performed
much faster than moving the stage.
We have developed a "Point-and-Shoot"
targeting system for microbeam irradiation based on the wide-field
magnetic split-coil deflector system from Technisches Büro Fischer (Ober
Ramstadt, Germany). The deflector consists of four interlaced coils in a
set of ferrite rings. The coils are paired to create two axes that are
orthogonal to each other. When wired in series, a pair of coils on
opposite sides will create a temporary dipole magnetic field, which
deflects the beam in the direction of the right hand rule. This system
has been used for the microbeam facility at Gesellschat für
Schwerionenforschung (GSI), Darmstadt, Germany. The deflector can move
the beam to as many as 1000 separate locations per second (more than 5
times the speed of movement of the stage!), which will dramatically
reduce the irradiation time.
It is possible to "scan" the beam into
a rectangular area in the microscope view by using the two pairs of
coils independently. The current needed in each coil to deflect the beam
to a specific location is mapped and this mapping used as a look-up
table for the target locations. The irradiation then proceeds by rapidly
switching through the required sets of coil currents to steer the beam
to the targets, irradiating each target with the time limited only by
the beam current.
We have installed Point-and-Shoot
systems on both the Permanent Magnet Microbeam (PMM) and the
electrostatic microbeam.
X-Ray Microbeam
Design
Nearly all microbeam facilities
currently employed for radiobiological applications use charged
particles - from protons to heavy ions, with LETs (stopping powers)
ranging from a few tenths to several hundred keV/µm. There are however
considerable benefits in using soft x-ray microbeams for both
mechanistic and risk estimation end-points. The higher spatial
resolution achievable with modern state-of-the-art x-ray optics elements
combined with the localized damage produced by the absorption of low
energy photons (~1 keV) represents a unique tool to investigate the
radio-sensitivity of sub-cellular and eventually sub-nuclear targets.
Moreover, as low-energy x rays undergo very little scattering, by using
x rays with an energy of ~5 keV it will be possible to irradiate with
micron precision individual cells and/or parts of cells up to a few
hundred microns deep inside a tissue sample in order to investigate the
relevance of effects such as the bystander effect in 3-D structured cell
systems.
We are upgrading the RARAF microbeam to include soft x rays:
characteristic Kα x rays from Ti, 4.5 keV (higher energies
are not feasible due to Compton scattering effects). The x-ray microbeam
(left) is mounted on the end of a horizontal beam line on the first
floor of RARAF. Because the x rays are being produced by reflection
instead of transmission, the x-ray beam will be vertical.
The x rays will be generated using an
electrostatic quadrupole quadruplet lens system to focus protons onto a
thick Ti target (best cross sections is at 4.5 MeV). The target consists
of a small plug of Ti pressed into a water-cooled copper block. The x
rays generated are demagnified using a zone plate. By using the already
focused proton microbeam to generate characteristic x rays, it is
possible to obtain a nearly monochromatic x-ray beam (very low
bremsstrahlung yield) and a reasonably small x-ray source (~20 µm
diameter), reducing the requirements on the zone plate.
| |
 |
| |
SEM micrograph
of our Fresnel zone plate |
Based on these parameters our zone plate specifications were
determined. The zone plate has a radius of 120 µm and an outmost zone
width of 50 nm. Preliminary off-line tests confirmed the focusing
characteristics of the zone plate. The zone plate has been placed
relatively close to the x-ray source (250 mm) and has a focal length of
23 mm (demagnification factor of ~11). Testing of the x-ray microbeam
system is progressing. The final expected dose rate to the sample for a
proton beam with a current of 10 nA will be about 30 cGy/min delivered
in a 2 micron spot.
Sub-Cellular Targeting
Typical
examples of sub-cellular targets for the RARAF microbeam have been: 1)
the cell nucleus or 2) the cell cytoplasm. Now that a focused
charged-particle beam with a sub-micron diameter is routinely available
using the compound electrostatic quadrupole triplet lens on the RARAF
microbeam, additional targets within cellular systems are accessible.
For instance, preliminary radiation experiments that target mitochondria
have been conducted on small airway epithelial cells. In these
experiments, the multiphoton microscope was used to image GFP-tagged
mitochondria sites and position them over the ion beam for irradiation.
The figure shows a multiphoton microscope image of mitochondria (green)
targets within small airway epithelial cells. The crosshairs (red) mark
the center of the image, which coincides with the position of the ion
beam. For this image and for post-irradiation time-lapse images of
mitochondria behavior, Hoechst was used to counter stain the nuclei.
Beam
Spot Diameter
The diameter of the beam spot for the
electrostatic focusing system has continued to decrease with tuning of
the focusing systems and development of new lenses, as can be seen in
the table below.
The most recent development has been
the installation of a compound quadrupole triplet. Two identical
triplets 2.20 meters apart are organized in Russian symmetry. Their
combined demagnifications have produced our first sub-micron beam spot!
The process of adjusting the alignment of the lenses and tuning of the
lens voltages is continuing with the aim of obtaining a beam spot
diameter of 0.5 µm.
|
Years |
Type of
beam |
Diameter (µm) |
|
2001 |
Focused -
single quadrupole quadruplet |
11 |
|
2002-2003 |
Focused -
single quadrupole quadruplet |
3 |
|
2004-2005 |
Focused -
single quadrupole triplet |
3.5 |
|
2006 |
Focused -
single quadrupole triplet |
2.0 |
|
2007 |
Focused -
single quadrupole triplet |
1.3 |
|
2008 |
Focused -
compound quadrupole triplet |
0.8 |
Multiphoton Microscope
Multiphoton microscopy, a laser-based,
3D imaging technique is now integrated into the Microbeam II endstation
at RARAF. The multiphoton microscope was custom-designed around the
Nikon Eclipse E600-FN research fluorescence microscope at the endstation
and is intended for imaging cell dynamics in tissue and cell-culture
samples following irradiation. Intended for detecting and observing
short-term molecular kinetics of radiation responses in living tissue
and in cell-culture samples, this multiphoton microscope is the first of
its kind to be assembled onto a microbeam cell-irradiation platform.
Multiphoton excitation is a phenomenon
where multiple photons, coincident in time and space, can act like a
single photon having a superposition of energies from the incident
photons. With two photon excitation, for example, two incident photons
at a laser focal point can act like one photon of half the wavelength. A
Chameleon Ultra II (Coherent Inc.) with a 680-1080 nm wavelength range
is the light source for the multiphoton microscope at RARAF. These
infrared wavelengths are less photo-damaging and have greater
penetration depth than single-photon excitation. At this laser's focal
point, the two-photon process effectively provides 340-540 nm. These
characteristics support exciting fluorophores, including RFP, at the
laser focal plane, and when the imaging technique is coupled to a
precision xyz-stage, optical sectioning in 3D tissue samples is
possible.
 |
|
|
Figure 1.
Multiphoton microscope at RARAF for observing responses in cell
cultures and in tissues after particle irradiation. |
|
Pictured in Figure 1, the setup of the
multiphoton microscope at RARAF includes a scanning laser source that is
built into the endstation microscope. The laser path starts at the Ti:S
laser, which is bolted to the optics table. The light path proceeds
through a variable attenuator and a fast shutter before being directed
vertically along the rotation axis of the microscope's pivot mount,
which is used to switch between online and offline positions. A beam
expander increases the beam to a size that will over-fill the back
aperture of the objective lens. The platform attached to the microscope
mount accommodates the scanning optics. These optics components point
along an axis into the side of a custom-built extension tube for the CCD
camera that is used during fluorescence microscopy. A pivoting mirror
inside the extension tube selects between fluorescence microscopy or
multiphoton microscopy. For multiphoton microscopy, this mirror guides
the laser vertically down through both the microscope tube lens and
objective lens.
Fluorophores in a sample absorb the
laser light through multiphoton excitation at the laser focal point. As
these markers fluoresce, signal light entering the objective is
deflected towards photomultiplier detectors (H5783, Hamamatsu). Shown in
figure 2 (below), signal detection is accomplished with photomultiplier
tubes that are mounted to the side of the filter cube changer on the
microscope. Control software monitors the correlation between the laser
scan position and signal at the photomultiplier tube to produce the
images.
An emission filter placed before a
photomultiplier tube designates the wavelength range (window) of
detected signal. With multiple fluorophores, color images can be
produced using appropriate dichroic mirrors and emission filters. In
figure 3 (below), the image of bovine pulmonary artery endothelial cells
(FluoCells prepared slide #1, Molecular Probes) is an example of a color
image produced by wavelength-specific detection. The cells in this image
were stained with BODIPY FL phallacidin to label the filamentous actin
(F-actin) and were counterstained with DAPI to label the nucleus. The
objective lens used for imaging these cells was a Nikon CFI LU Epi Plan
Fluor 50X objective, with 0.80 numerical aperture and 1.0 mm working
distance. To develop this image, 20 to 30 single frames were acquired
per color at a rate of 2 seconds/frame. A routine written in MATLAB®
subsequently summed and combined the images into a 3D matrix for color
display.
|
|
 |
|
 |
|
Figure 2.
Fluorescent signal emitted by the sample is deflected by a dichroic
mirror and collected by photomultiplier tubes |
Figure 3.
Multiphoton microscopy image of bovine pulmonary artery endothelial
cells. (FluoCells prepared slide #1, Molecular Probes). Cell nuclei
diameter ~10 microns. |
With the attachment housing for two
photomultiplier tubes, experiments using fluorescence resonance energy
transfer (FRET) are possible. FRET optics sets are on hand for the
following transitions: BFP > GFP, CFP > YFP, and Alexa 488 > Alexa 594.
FRET experiments concurrent with cell irradiation are examples of
monitoring post cell-irradiation dynamics using the multiphoton
microscope at RARAF.
Particle-Induced Foci Formation
One of our collaborators (David Chen,
University of Texas Southwestern, Dallas, TX is using the microbeam to
examine particle-induced single strand breaks in DNA within single
cells. They are using human HT1080 fibrosarcoma cells that have been
transfected with a GFP-tagged XRCC1 DNA single-strand break repair
protein. Figure 2 shows multiphoton images of one cell a) before
irradiation and b) 4 minutes after exposure to 400 3-MeV alpha particles
at a predetermined position within a cell nucleus. Four minutes after
irradiation, the region of enhanced GFP signal corresponds to the XRCC1
focus and the irradiation position. In this example, ion-beam targeting
was deliberate to avoid areas deficient in GFP (likely nucleoli). This
clear response demonstrates that the multiphoton microscope was
successful at recording high-resolution, time-lapse images of
particle-induced focus formation within a single cell in real time. As
an additional capability, projections of 3D z-stacks can quantify focus
track volume to correlate with the amount of radiation (number of
particles) delivered to the cell nucleus.
The system is fully operational.
For more detail see:
Microbeam-integrated multiphoton imaging system. Bigelow AW, Geard
CR, Randers-Pehrson G, and Brenner DJ. Rev. Sci. Instruments 79:123707
(2008).
Permanent Magnet
Microbeam
The
permanent magnet microbeam (PMM), under development at RARAF, presents
an alternative approach to microbeam design. Instead of focusing the ion
beam using electromagnetic or electrostatic lenses, this system uses
permanent magnets, which require no power supplies.
4He ions from the
accelerator are focused using a compound magnetic lens consisting of two
quadrupole triplets. The first triplet is placed 2 m above the object
aperture, with a second (identical) triplet placed 2 m above the focal
plane of the first. Since each triplet does not have identical
demagnifications in the x and y axes, the two lenses are rotated by 90°
in the x-y plane so that a circular beam spot is obtained (Russian
Symmetry). The object aperture was initially covered with a 1.8-μm thick
Al scattering foil (since replaced by a phase space sweeper), used to
eliminate any correspondence between angle and position for the
particles in the beam. A limiting aperture is placed before the first
triplet and inside the second triplet to reject ions which have very
large aberrations.
The cells to be irradiated are placed
at the image plane of the compound lens. The PMM is mounted on the
original microbeam endstation consisting of a microscope with a particle
detector mounted on one of the objective lenses and an x-y stage
positioned by stepping motors.
Magnetostatic Lens
The
design of the compound lens, used to focus the beam, is based on the one
used for our electrostaticly focused microbeam. However, in order to
simplify the PMM operation we have elected to use permanent magnets to
construct the lens as opposed to the electrostatic lenses. The use of
permanent magnets eliminates the need for bulky power supplies and
cooling systems required by other types of ion lenses in addition to
allowing a tighter configuration (and therefore better optical
properties) than common electromagnetic lenses. Magnet strength is
adjusted by moving rare earth magnets in and out of a shaped yoke as
seen in the figure.
Two permanent magnet quadrupole
triplets have been purchased from
STI
Optronics. The optimized lens, shown here, consists of two outer
4.25 cm long magnetic quadrupoles and an 8.5 cm long center quadrupole
with inter-quadrupole gaps of 1.67 cm and a bore of radius 6.35 mm. It
should be noted that such a small bore radius is rather difficult to
obtain with standard electromagnets.
Beam Tests
Just prior to the
decommissioning of the RARAF Van de Graaff accelerator in June 2005, we
had attained a beam spot size of 20 µm. As a first step, the beam was
imaged at the focal plane of the first quadrupole triplet, using a
commercial CCD chip. The observed spot size of 50 x 150 µm was in good
agreement with that expected from simulations.
As a second step,
the beam diameter was measured at the endstation using the knife edge
technique. The spot size was tuned by adjusting all magnets while
maintaining Russian symmetry - in particular we tried to keep
quadrupoles 1, 3, 4 and 6 at the same strength (A) and quadrupoles 2 and
5 at the same strength (B). The figure shows the theoretical and
measured spot size and shape at the end station. While the general
trends are very similar in both cases, the smallest spot size obtained
experimentally was only 20 µm in diameter, two times larger than the
theoretical prediction. Simulations have shown that this is probably due
to residual high-order fields in the quadrupoles or due to misalignment.
In 2006, the magnetic quadrupole system had to be
removed for the construction of the laboratories on the third floor and
was reassembled in early 2007. Without adjusting the magnets,
we measured a beam spot size of 20 microns, demonstrating the robustness
of this design. Following the replacement of the scattering foil with a
phase space sweeper and smaller aperture we have obtained a beam spot
approximately 8 µm in diameter.
The magnetostatic lens is in routine use.
For more details see:
Testing the stand-alone
microbeam at Columbia University. Garty G., Ross G.J., Bigelow A.,
Randers-Pehrson G. and Brenner D.J. Radiat. Prot. Dosim.
122:292-296, 2006.
A single-particle /
single-cell microbeam based on an isotopic alpha source.
Ross G.J., et al. Nucl. Instrum. Meth.B. 231:207-211, 2005.
A microbeam
irradiator without an accelerator. Garty G.,
et al. Nucl. Instrum. Meth. B 241:392-396, 2005.
Immersion Mirau
Interferometry
An
immersion-based Mirau interferometric (IMI) objective has been designed
and built to function as an immersion lens with standard interferometric
techniques using a short coherence length and to otherwise accommodate
the endstation requirements for the Columbia University microbeam at
RARAF.
To enable interferometry, we used high
precision glass for the beamsplitter and the mirror. The surfaces
interacting with the medium have a protective layer on top of the
coating.
The new attachment has been tested as
a water immersion lens. We confirmed that the two equal-arm light
pathways are indeed restored and then provide interference fringes in
the environment with sufficient contrast to perform the biological
experiments. The acquired images are comparable to those that can be
obtained using the commercial attachment that we used for preliminary
testing.
|
|
 |
|
Diagram of the
immersion-based Mirau lens |
The new objective will support rapidly
and automatically locating the cell nuclei without stain. The "banding"
in some of the examples shows the presence of both a slow drift in
z-position and some very slight vibrational interference, due to the
requirement that interference patterns needed to reproduce the image are
acquired at different times. Even very low amplitude vibrations, if they
exist, may prevent maintaining the desirable phase shifts between
consecutive frames. One solution is to acquire all interference images
at the same time, presently being
developed at RARAF.
|
|
 |
|
 |
|
Custom-built
interferometric attachment with an additional window unit |
Left: IMI
image of normal human dermal fibroblasts taken with the interim
immersion-Mirau lens and a neutral density microscope cube. Right:
Same sample with a UV cube to reveal the nuclear stain. |
For more details see:
Phase-based cell imaging
techniques for microbeam irradiations. G.J.
Ross , et al. Nucl Instr Meth B
241: 387-391 (2005)
Laser Ion Source
The standard RF ion source that is
used on our 5 MV HVE Singletron particle accelerator ionizes atoms from
the gas phase and is suitable for experiments involving the irradiation
of cells with protons, deuterons and alpha particles having linear
energy transfers (LETs) from 10 to 200 keV/μm. To extend the upper LET
range of our experiments, ions as heavy as iron (Table 1) are required
with high enough energies to have usable particle ranges ≥ 20 μm.
Because we are limited to a terminal voltage of 5 MV, highly charged
initial heavy ions are needed to obtain the required energies. Laser ion
sources, in which a high-power, pulsed light beam is focused to micron
dimensions, resulting in power densities on the order of 1012
W/cm2, can produce high initial charge-states of heavy ions
in sufficient fluxes for microbeam experiments.
Table 1.
Potential Ion Beams for Microbeam Irradiations
|
Ion |
Atomic Number |
Mass
(amu) |
Ion Energy
(MeV) |
Min. Charge State for Acceleration |
Terminal Potential
(MV) |
Stopping Power
(keV/μm) |
Range
(μm) |
|
p |
1 |
1 |
5.00 |
1 |
5.00 |
8.4 |
340 |
|
d |
1 |
2 |
1.20 |
1 |
1.20 |
40 |
20 |
|
He |
2 |
4 |
10.00 |
2 |
5.00 |
59 |
105 |
|
Li |
3 |
7 |
15.00 |
3 |
5.00 |
148 |
68 |
|
Be |
4 |
9 |
15.00 |
3 |
5.00 |
310 |
37 |
|
B |
5 |
11 |
15.00 |
3 |
5.00 |
515 |
25 |
|
C |
6 |
12 |
20.00 |
4 |
5.00 |
605 |
28 |
|
N |
7 |
14 |
20.00 |
4 |
5.00 |
845 |
22 |
|
O |
8 |
16 |
25.00 |
5 |
5.00 |
955 |
24 |
|
Al |
13 |
27 |
40.00 |
8 |
5.00 |
1960 |
22 |
|
Cl |
17 |
35 |
55.00 |
11 |
5.00 |
2760 |
22 |
|
Ti |
22 |
48 |
70.00 |
14 |
5.00 |
3925 |
23 |
|
Fe |
26 |
56 |
75.00 |
15 |
5.00 |
4780 |
22 |
Our
laser ion source (LIS) is based on laser ablation. High-power Nd:YAG
laser pulses can enter the accelerator tank through a window, pass
through the SF6 insulating gas, and enter a vacuum region
through an optical port to deliver a focused, high-power-density pulse
onto a solid target. Ions created during an ablation event’s energy
transfer are ejected in the form of a plasma plume in a direction that
is preferentially normal to the solid target surface. These ions are
emitted with a distribution in charge state, kinetic energy, and angle.
Following a drift distance for plasma expansion, a spherical
electrostatic analyzer (ESA) selects ions with a particular energy per
charge and focuses them to a point coinciding with the 3.18-mm entrance
aperture of the accelerator tube. The LIS is pictured in a table-top
configuration left.
To maximize the ion flux, a Quanta-Ray
LAB-190-100 Nd:YAG pulsed laser system from Spectra Physics was chosen
for its high repetition rate (100 Hz) and high power (325 mJ/pulse, 10
ns Q-switch pulse duration). A focusing lens (f = 12.2 cm) for the laser
is mounted on a holder inside the vacuum system and the focal spot
diameter (61 μm) ideally provides a focused power density of 2.2 x 1012
W/cm2. A 2 mm thick piece of AR-coated, BK7 glass is used as
a lens protector to inhibit target spatter from accumulating on the
focusing lens. The ablation target is a cylindrical solid attached to a
differential drive manipulator. The assembly drawing for the Columbia
University LIS is shown below.
Ions
generated from the laser ablation process have a wide distribution in
charge state and energy. To select a particular ion charge state and
energy, the plasma first expands through a drift distance of 70 cm and
then enters a 24-degree spherical ESA that was designed and built
in-house. This ESA is a double-focusing element with fringe field shunts
and interchangeable entrance and exit apertures. A photograph of our
spherical ESA is shown below. The ESA is a self-contained module mounted
on a six-inch flange. An electrostatic field within the ESA selects ions
by energy/charge. So for each pulsed, laser-ablation event, the ions
that pass through the ESA can be characterized by a time-of-flight (TOF)
spectrum. The highest charge-state ions have the highest energy and
arrive at the accelerator entrance aperture first, while the lowest
charge-state ions arrive last.
Several
considerations were made to tailor-fit the LIS footprint into the
Singletron machine. Auxiliary power available at the Singletron terminal
is limited, but adequate, for a non-evaporable getterer (NEG) pump to
maintain proper vacuum conditions within the target ablation area.
Additionally, it is crucial to have a match between the ion source
emittance and the acceptance of the Singletron accelerator tube. In
collaboration with HVE, appropriate modifications were made to the
primary accelerator tube sections to optimize the transmission of the
LIS-produced ions and also to conserve the function of the primary RF
ion source incorporated on the Singletron.
For more details see:
Laser ion source for the
Columbia University microbeam.
(Nucl. Instrum. Meth. B
241:874-879, 2005)
Secondary
Electron Ion Microscope
As we improve the spatial
characteristics of the microbeam system, it becomes increasingly
important to be able to assess the beam quality, in order to adjust the
system to its optimum capabilities. For this purpose we are developing a
secondary electron ion microscope (SEIM). This device will enable us to
measure the beam profile and position, in real time, with sub-micron
resolution.
The
SEIM design is inspired by the common technique of photoelectron
microscopy (PEM). It is based on conversion of incident projectiles
using a secondary electron emitting (SEE) film, generating one (or more)
electrons per projectile. The ejected electrons are then imaged, using
an electrostatic unipotential lens, forming a 500 times magnified image
on an image intensified CCD This high magnification enables 100 nm
resolution using a 50 μm resolution CCD.
In order to overcome the chromatic and
spherical aberrations inherent in the electrostatic lens, the electrons
are bent by a 45ş angle, reflected by an electrostatic mirror and bent
by an additional 45ş before reaching the detector. This “folded” design
of the SEIM is a novel one, developed at RARAF.
From
systematic simulations we have found the “ideal” SEIM configuration (a
150° mirror biased at 1.5 V with respect to the SEE film as well as the
required lens voltage as a function of mirror placement), giving a
single-electron resolution of 300-500 nm and single electron transport
efficiency of 55%-80%, depending on the mirror bias. This should be
compared with ~1% for “straight” geometries utilizing an aperture to
eliminate energetic electrons). We expect that this resolution will
improve as the square root of the electron yield which is expected to be
higher than unity for heavy/energetic ions, where the SEIM is needed
most.
Preliminary tests, using an “unfolded”
SEIM, consisting of the electrostatic lens and the electron detector but
without the magnet and mirror agree well with simulations. For these
measurements we replaced the SEE foil with a quartz window on which a
micron scale pattern of aluminum was evaporated. The pattern was
illuminated with low intensity UV light and the resulting photoelectrons
were imaged. A comparison of the optical- and electron-based images of
the pattern are shown on the right. The width of the spot edge allows us
to estimate the SEIM resolution at 4.3 μm RMS. This is in good agreement
with the simulated prediction of 4-5 μm. The predicted magnification
(16x) is also in good agreement with the measured 20x.
For more details see:
Development of a secondary-electron ion-microscope for microbeam
diagnostics. Garty G., Randers-Pehrson G. and Brenner D.J. Nucl.
Instr. Meth. B. 231:60-64 (2005).
|