Our current single-particle microbeam
system features an extensively upgraded spatial resolution of 0.8 μm
(diameter) for the beam. Spatial resolution has been improved by use of our focused
lens system and with the aid of our new
accelerator.
Installation of our laser
ion source will provide more densely-ionizing
radiations, i.e., heavier ion beams, with stopping powers from 10 to 4,500 keV/μm,
compared with our current beams of H and He ions, which have stopping powers
from
10 to 200 keV/μm. A secondary
electron ion microscope is available to assess beam quality. New
state-of-the-art facilities are available for biology.
The diameter of the beam spot for the electrostatic focusing system has continued to decrease with tuning of the focusing systems and development of new lenses, as can be seen in the table
below.
The most recent development has been the installation of a compound quadrupole triplet. Two identical triplets 2.20 meters apart are organized in Russian symmetry. Their combined demagnifications have produced our first sub-micron beam spot! The process of adjusting the alignment of the lenses and tuning of the lens voltages is continuing with the aim of obtaining a beam spot diameter of 0.5 µm.
Multiphoton microscopy, a laser-based, 3D imaging technique is now integrated into the Microbeam II endstation at RARAF. The multiphoton microscope was custom-designed around the Nikon Eclipse E600-FN research fluorescence microscope at the endstation and is intended for imaging cell dynamics in tissue and cell-culture samples following irradiation.
Intended for detecting and observing short-term molecular kinetics of
radiation responses in living tissue and in cell-culture samples, this multiphoton microscope is the first of its kind to be assembled onto a microbeam cell-irradiation platform.
Multiphoton excitation is a phenomenon
where multiple photons, coincident in time and space, can act like a
single photon having a superposition of energies from the incident
photons. With two photon excitation, for example, two incident photons at
a laser focal point can act like one photon of half the wavelength. A Chameleon
Ultra II (Coherent Inc.) with a 680-1080 nm wavelength range is the light source for the multiphoton microscope at RARAF.
These infrared wavelengths are less photo-damaging and have greater penetration depth than single-photon excitation.
At this laser's focal point, the two-photon process effectively provides
340-540 nm. These characteristics support exciting fluorophores,
including RFP, at the laser focal plane, and when the imaging technique is coupled to a precision xyz-stage, optical sectioning in 3D tissue samples is possible.
Figure 1. Multiphoton microscope at RARAF for observing responses in cell cultures and in tissues after particle irradiation.
Pictured in Figure 1, the setup of the multiphoton microscope at RARAF includes a scanning laser source that is built into the endstation microscope. The laser path starts at the Ti:S laser, which is bolted to the optics table. The light path proceeds through a variable attenuator and a fast shutter before being directed vertically along the rotation axis of the microscope's pivot mount, which is used to switch between online and offline positions. A beam expander increases the beam to a size that will over-fill the back aperture of the objective lens. The platform attached to the microscope mount accommodates the scanning optics. These optics components point along an axis into the side of a custom-built extension tube for the CCD camera that is used during fluorescence microscopy. A pivoting mirror inside the extension tube selects between fluorescence microscopy or multiphoton microscopy. For multiphoton microscopy, this mirror guides the laser vertically down through both the microscope tube lens and objective lens.
Fluorophores in a sample absorb the laser light through multiphoton excitation at the laser focal point. As these markers fluoresce, signal light entering the objective is deflected towards photomultiplier detectors (H5783, Hamamatsu). Shown in figure 2
(below), signal detection is accomplished with photomultiplier tubes that are mounted to the side of the filter cube changer on the microscope. Control software monitors the correlation between the laser scan position and signal at the photomultiplier tube to produce the images.
An emission filter placed before a photomultiplier
tube designates the wavelength range (window) of detected signal. With
multiple fluorophores, color images can be produced using appropriate
dichroic mirrors and emission filters. In figure 3 (below), the image of bovine
pulmonary artery endothelial cells (FluoCells prepared slide #1, Molecular
Probes) is an example of a color image produced by wavelength-specific
detection. The cells in this image were stained with BODIPY FL phallacidin
to label the filamentous actin (F-actin) and were counterstained with
DAPI to label the nucleus. The objective lens used for imaging these
cells was a Nikon CFI LU Epi Plan Fluor 50X objective, with 0.80 numerical
aperture and 1.0 mm working distance. To develop this image, 20 to 30
single frames were acquired per color at a rate of 2 seconds/frame.
A routine written in MATLAB® subsequently summed and combined
the images into a 3D matrix for color display.
Figure 2. Fluorescent signal emitted by the sample is deflected by a dichroic mirror and collected by photomultiplier tubes
With the attachment housing for two photomultiplier tubes, experiments using fluorescence resonance energy transfer (FRET) are possible. FRET optics sets are on hand for the following transitions: BFP > GFP, CFP > YFP, and Alexa 488 > Alexa 594. FRET experiments concurrent with cell irradiation are examples of monitoring post cell-irradiation dynamics using the multiphoton microscope at RARAF.
Particle-Induced Foci Formation
One of our collaborators (David Chen,
University of Texas Southwestern, Dallas, TX is using the microbeam to
examine particle-induced single strand breaks in DNA within single
cells. They are using human HT1080 fibrosarcoma cells that have been
transfected with a GFP-tagged XRCC1 DNA single-strand break repair
protein. Figure 2 shows multiphoton images of one cell a) before
irradiation and b) 4 minutes after exposure to 400 3-MeV alpha particles
at a predetermined position within a cell nucleus. Four minutes after
irradiation, the region of enhanced GFP signal corresponds to the XRCC1
focus and the irradiation position. In this example, ion-beam targeting
was deliberate to avoid areas deficient in GFP (likely nucleoli). This
clear response demonstrates that the multiphoton microscope was
successful at recording high-resolution, time-lapse images of
particle-induced focus formation within a single cell in real time. As
an additional capability, projections of 3D z-stacks can quantify focus
track volume to correlate with the amount of radiation (number of
particles) delivered to the cell nucleus.
The system is now fully operational.
As an example of the use of our multiphoton microscope...
The
permanent magnet microbeam (PMM), under development at RARAF, presents
an alternative approach to microbeam design. Instead of focusing the
ion beam using electromagnetic or electrostatic lenses, this system
uses permanent magnets, which require no power supplies.
4He ions from the accelerator
are focused using a compound magnetic lens consisting of two quadrupole
triplets. The first triplet is placed 2 m above the object aperture,
with a second (identical) triplet placed 2 m above the focal plane of
the first. Since each triplet does not have identical demagnifications
in the x and y axes, the two lenses are rotated by 90° in the x-y plane
so that a circular beam spot is obtained (Russian Symmetry). The object
aperture was initially covered with a 1.8-μm thick Al scattering
foil (since replaced by a phase space sweeper), used to eliminate any correspondence between angle and position
for the particles in the beam. A limiting aperture is placed before
the first triplet and inside the second triplet to reject ions which
have very large aberrations.
The cells to be irradiated are placed
at the image plane of the compound lens. The PMM is mounted on the original
microbeam endstation consisting of a microscope with a particle detector
mounted on one of the objective lenses and an x-y stage positioned by
stepping motors.
Magnetostatic Lens
The design of the compound lens, used to focus the beam, is based on the one used for our electrostaticly focused microbeam. However, in order to simplify the PMM operation we have elected to use permanent magnets to construct the lens as opposed to the electrostatic lenses. The use of permanent magnets eliminates the need for bulky power supplies and cooling systems required by other types of ion lenses in addition to allowing a tighter configuration (and therefore better optical properties) than common electromagnetic lenses. Magnet strength is adjusted by moving rare earth magnets in and out of a shaped yoke as seen in the figure.
Two permanent magnet quadrupole triplets
have been purchased from STI
Optronics. The optimized lens, shown here, consists of two outer
4.25 cm long magnetic quadrupoles and an 8.5 cm long center quadrupole
with inter-quadrupole gaps of 1.67 cm and a bore of radius 6.35 mm.
It should be noted that such a small bore radius is rather difficult
to obtain with standard electromagnets.
Beam Tests
Just
prior to the decommissioning of the RARAF Van de Graaff accelerator
in June 2005, we had attained a beam spot size of 20 µm. As a first
step, the beam was imaged at the focal plane of the first quadrupole
triplet, using a commercial CCD chip. The observed spot size of 50
x 150 µm was in good agreement with that expected from simulations.
As
a second step, the beam diameter was measured at the endstation using
the knife edge technique. The spot size was tuned by adjusting all magnets
while maintaining Russian symmetry - in particular we tried to keep
quadrupoles 1, 3, 4 and 6 at the same strength (A) and quadrupoles 2
and 5 at the same strength (B). The figure shows the theoretical and
measured spot size and shape at the end station. While the general trends
are very similar in both cases, the smallest spot size obtained experimentally
was only 20 µm in diameter, two times larger than the theoretical
prediction. Simulations have shown that this is probably due to residual
high-order fields in the quadrupoles or due to misalignment.
In 2006, the magnetic quadrupole system
had to be removed for the construction of the laboratories on the
third floor and was reassembled in early 2007. Without adjusting
the magnets, we measured a beam spot size of 20 microns,
demonstrating the robustness of this design. Following the replacement
of the scattering foil with a phase space sweeper and smaller aperture
we have obtained a beam spot reliably under 5µm. The PMM is
now available as an alternative to the electrostatic microbeam.
Immersion Mirau Interferometry (IMI) is a non-stain imaging technique that is
based on the principles of phase-shifting interferometry. Several
interference images are acquired at different heights with quarter-wavelength
phase shifts between their interference patterns. As an interferometric
technique, IMI does not require fluorescent staining of the cells,
which eliminates potential damage induced by fluorescent stains. It
operates at 540 nm (green light), therefore does not induce UV-exposure
of the cells. Lastly, immersion mode allows the experimenter to keep
the cells in cell medium during the irradiation experiment and minimize
cell death due to dehydration.
To facilitate the immersion operation
mode, a custom Mirau interferometric attachment is required. Such
attachment has been designed and built at RARAF to function as an
immersion lens with standard interferometric techniques using a short
coherence length.
Custom-built interferometric attachment. The shape is designed
to fit the microbeam dish. Two adjustment screws regulate the
inclination of the optics with respect to the sample. Aluminum
surface is anodized to protect corrosion by the medium. The
optics holders are interchangeable. Also attached to the objective
is the holder for capacitive sensor.
To enable interferometry, we used high precision glass
for the beamsplitter and the mirror. The surfaces interacting with
the medium have a protective layer on top of the coating.
IMI image of unstained living fibroblasts.
The new attachment has been tested as a water immersion
lens. We confirmed that the two equal-arm light pathways are indeed
restored and then provide interference fringes in the environment
with sufficient contrast to perform the biological experiments. The
acquired images are comparable to those that can be obtained using
the commercial attachment that we used for preliminary testing.
In case of interferometric imaging,
even a fraction-of-a-wavelength shift in the vertical direction lowers
the image quality. Thus, despite the fact that the microbeam is mounted
on the 27" thick concrete floor, the IMI system is still sensitive
to unpredictable low-amplitude vibrations.
A recent development is Simultaneous Mirau Interferometry (SIMI). SIMI is a vibration-independent version of IMI and has shorter image acquisition time. SIMI is an integration of Immersion Mirau Interferometry and polarization microscopy.
Quantitative Phase Imaging
Software-based
Quantitative Phase Imaging (QPI) is now available for use in the Columbia
microbeam laboratory as a non-interferometric approach to no-stain
imaging. Reflected-light based images are obtained in focus and with
the focus set slightly above and below the sample plane. These images
are then used to approximately solve the light transport equation
using the Fourier transform-based software from Iatia (Melbourne,
Australia), a set of relatively new techniques that generate phase
images and phase-amplitude images. The results are used to create
a new 2-D map of the sample which is then used by the custom microbeam
irradiation software to locate cells.
In some of the tests, some cells have
been missed and there also have been false positives. In general,
the quality of the images can be improved by careful tuning of the
parameters for the approximate solution to the transport equation.
However, to optimize the system for our regular, automated use, there
is still work to be done in eliminating false cells and reducing missed
cells. Several variables have been isolated and eliminated as causes:
plating time, cell-type, cell phase, light color, cell growth surface,
amount of medium (depth), percent of medium vs. buffer, and use of
a cover glass.
In the picture the QPI algorithm has
been applied to normal human dermal fibroblasts under conditions conducive
to rapid automatic cell location.
QPI and IMI have been evaluated in
comparison to each other. Under consideration were: processing time,
reliability, maintenance and ease of use for the experimenter. QPI
is a software solution and does not require introduction of fluids
or cleaning, whereas IMI does. QPI does not require any additional
equipment, while the immersion-Mirau interferometer requires the end-user
to use a custom objective. IMI is sensitive to vibrations unlike QPI.
At the same time, QPI is more sensitive to the algorithm parameters,
which makes Immersion-Mirau interferometry more reliable and ready
to automate than QPI. Work is currently being done on vibration-independent approach for
IMI.
The new biology lab as viewed from the
entrance to the third floor laboratory space (upper right in layout
figure). In view are the two islands that have work stations for up to
12 people.
In 2006, the Trustees of Columbia University contributed
the funds ($1.8M) required to build over 2000 square feet of new
laboratory and office space on the third floor of the facility. When RARAF
was built in the early 1980s, this space had been intended for offices and
a meeting room. Since there were never funds available to build these, the
space had been used for storage. Construction began in early April 2006
and is now completed. Over half the area is a biology laboratory with
a class II laminar flow hood (and provision for two additional laminar flow hoods),
several incubators, refrigerators and freezers, and two islands comprising
a dozen work stations. There are also six desks for technicians and postdocs.
In addition there is a physics laboratory and a dedicated microscopy
laboratory with three workstations.
The layout of the
third floor of RARAF. Microbeam lab III (permanent magnet lens
system) is directly above the original microbeam laboratory. The
wall on the left is ~58 feet long. Microbeam II, which houses
the electrostatic lens system, is shown in the bottom, right of
the center and is half a floor below the new construction, shown
with emphasized walls.
Also included in the construction is a new
laboratory for the permanent magnet microbeam facility. While this microbeam
system was being developed, it was housed in a makeshift room. The magnetic
quadrupole triplets and endstation for the system were removed before
construction began. Now that construction is complete, the beam line
has been reconstructed with the two triplet lenses, and the endstation
with the microscope and electronics has been installed. The lab is fitted
with a laminar flow hood and an incubator. The magnet system is being
adjusted to provide the best focusing using proton and helium ion beams
from the accelerator. When the focusing studies are complete, the system
will be available again for biological irradiations.
RARAF
was originally based on a Van de Graaff linear particle accelerator
that by 2005 was over 55 years old. In order to stay on the cutting
edge, we must focus our particle beams to even smaller sub-cellular
dimensions. This requires an energy stability that the original
accelerator simply could not produce but that is achieved with the
newest generation of linear accelerators.
A 5 MV Singletron (right)
purchased from High Voltage Engineering Europa (HVEE) in the Netherlands
has replaced
our Van de Graaff. Installation of the Singletron was completed at the
end of 2005. The main technical advantage of this
new accelerator is the amount of voltage fluctuation. Energy
fluctuations cause an effect similar to wavelength changes in optical
lenses because of “chromatic” aberrations in the lenses. The lower the
energy fluctuation, the better the electrostatic lens system is able to focus the beam.
Our Van de Graaff used a cloth belt, and charge was carried mechanically
to the terminal. The Singletron has no moving parts; the high voltage
is generated electronically by a Cockcroft-Walton type power supply
and therefore is more stable. It is easier to compensate for electronic
fluctuations than mechanical fluctuations. A voltage ripple of 200 V
peak-to-peak at 3.75 MV is guaranteed with a design aim of 50 V.
A terminal potential of 5 MV also means that
using the laser
ion source we can obtain
heavy ions with usable ranges using lower charge states or longer ranges (higher energies) at the same charge state.
It also allows us to obtain higher-energy proton beams so that LETs
below 10 keV/μm are available for low-LET experiments on both
the microbeam and the track segment (broad beam) facilities.
The new accelerator has been running biological experiments for over two years. The low fluctuation in terminal voltage has helped in the reduction of the beam spot sizes for the PMM and, especially, the electrostatically focused microbeam. The higher terminal voltage has provided more penetrating beams
of He ions and protons, which have been used for irradiation of tissue samples using the Track Segment Facility.
The standard RF ion source that is used
on our 5 MV HVE Singletron particle accelerator ionizes atoms from the gas
phase and is suitable for experiments involving the irradiation of cells
with protons, deuterons and alpha particles having linear energy transfers
(LETs) from 10 to 200 keV/μm. To extend the upper LET range of our
experiments, ions as heavy as iron (Table 1) are required with high enough
energies to have usable particle ranges ≥ 20 μm. Because we are limited to
a terminal voltage of 5 MV, highly charged initial heavy ions are needed
to obtain the required energies. Laser ion sources, in which a high-power,
pulsed light beam is focused to micron dimensions, resulting in power
densities on the order of 1012 W/cm2, can produce
high initial charge-states of heavy ions in sufficient fluxes for
microbeam experiments.
Table 1.
Potential Ion Beams for Microbeam Irradiations
Ion
Atomic Number
Mass
(amu)
Ion Energy
(MeV)
Min. Charge State for Acceleration
Terminal Potential
(MV)
Stopping Power
(keV/μm)
Range
(μm)
p
1
1
5.00
1
5.00
8.4
340
d
1
2
1.20
1
1.20
40
20
He
2
4
10.00
2
5.00
59
105
Li
3
7
15.00
3
5.00
148
68
Be
4
9
15.00
3
5.00
310
37
B
5
11
15.00
3
5.00
515
25
C
6
12
20.00
4
5.00
605
28
N
7
14
20.00
4
5.00
845
22
O
8
16
25.00
5
5.00
955
24
Al
13
27
40.00
8
5.00
1960
22
Cl
17
35
55.00
11
5.00
2760
22
Ti
22
48
70.00
14
5.00
3925
23
Fe
26
56
75.00
15
5.00
4780
22
Our
laser ion source (LIS) is based on laser ablation. High-power Nd:YAG
laser pulses can enter the accelerator tank through a window, pass
through the SF6 insulating gas, and enter a vacuum region
through an optical port to deliver a focused, high-power-density pulse
onto a solid target. Ions created during an ablation event’s energy
transfer are ejected in the form of a plasma plume in a direction
that is preferentially normal to the solid target surface. These ions
are emitted with a distribution in charge state, kinetic energy, and
angle. Following a drift distance for plasma expansion, a spherical
electrostatic analyzer (ESA) selects ions with a particular energy
per charge and focuses them to a point coinciding with the 3.18-mm
entrance aperture of the accelerator tube. The LIS is pictured in
a table-top configuration left.
To maximize the ion flux, a Quanta-Ray
LAB-190-100 Nd:YAG pulsed laser system from Spectra Physics was chosen for
its high repetition rate (100 Hz) and high power (325 mJ/pulse, 10 ns
Q-switch pulse duration). A focusing lens (f = 12.2 cm) for the laser is
mounted on a holder inside the vacuum system and the focal spot diameter
(61 μm) ideally provides a focused power density of 2.2 x 1012
W/cm2. A 2 mm thick piece of AR-coated, BK7 glass is used as a
lens protector to inhibit target spatter from accumulating on the focusing
lens. The ablation target is a cylindrical solid attached to a
differential drive manipulator. The assembly drawing for the Columbia
University LIS is shown below.
Ions
generated from the laser ablation process have a wide distribution in
charge state and energy. To select a particular ion charge state and
energy, the plasma first expands through a drift distance of 70 cm and
then enters a 24-degree spherical ESA that was designed and built in-house.
This ESA is a double-focusing element with fringe field shunts and interchangeable
entrance and exit apertures. A photograph of our spherical ESA is shown
below. The ESA is a self-contained module mounted on a six-inch flange.
An electrostatic field within the ESA selects ions by energy/charge.
So for each pulsed, laser-ablation event, the ions that pass through
the ESA can be characterized by a time-of-flight (TOF) spectrum. The
highest charge-state ions have the highest energy and arrive at the
accelerator entrance aperture first, while the lowest charge-state ions
arrive last.
Several
considerations were made to tailor-fit the LIS footprint into the Singletron
machine. Auxiliary power available at the Singletron terminal is limited,
but adequate, for a non-evaporable getterer (NEG) pump to maintain proper
vacuum conditions within the target ablation area. Additionally, it
is crucial to have a match between the ion source emittance and the
acceptance of the Singletron accelerator tube. In collaboration with
HVE, appropriate modifications were made to the primary accelerator
tube sections to optimize the transmission of the LIS-produced ions
and also to conserve the function of the primary RF ion source incorporated
on the Singletron.
As we improve the spatial
characteristics of the microbeam system, it becomes increasingly important
to be able to assess the beam quality, in order to adjust the system to
its optimum capabilities. For this purpose we are developing a secondary
electron ion microscope (SEIM). This device will enable us to measure the
beam profile and position, in real time, with sub-micron resolution.
The
SEIM design is inspired by the common technique of photoelectron microscopy
(PEM). It is based on conversion of incident projectiles using a secondary
electron emitting (SEE) film, generating one (or more) electrons per
projectile. The ejected electrons are then imaged, using an electrostatic
unipotential lens, forming a 500 times magnified image on an image intensified
CCD This high magnification enables 100 nm resolution using a 50 μm
resolution CCD.
In order to overcome the chromatic and
spherical aberrations inherent in the electrostatic lens, the electrons
are bent by a 45ş angle, reflected by an electrostatic mirror and bent by
an additional 45ş before reaching the detector. This “folded” design of
the SEIM is a novel one, developed at RARAF.
From
systematic simulations we have found the “ideal” SEIM configuration
(a 150° mirror biased at 1.5 V with respect to the SEE film as well
as the required lens voltage as a function of mirror placement), giving
a single-electron resolution of 300-500 nm and single electron transport
efficiency of 55%-80%, depending on the mirror bias. This should be
compared with ~1% for “straight” geometries utilizing an aperture to
eliminate energetic electrons). We expect that this resolution will
improve as the square root of the electron yield which is expected to
be higher than unity for heavy/energetic ions, where the SEIM is needed
most.
Preliminary tests, using an “unfolded”
SEIM, consisting of the electrostatic lens and the electron detector
but without the magnet and mirror agree well with simulations. For these
measurements we replaced the SEE foil with a quartz window on which
a micron scale pattern of aluminum was evaporated. The pattern was illuminated
with low intensity UV light and the resulting photoelectrons were imaged.
A comparison of the optical- and electron-based images of the pattern
are shown on the right. The width of the spot edge allows us to estimate
the SEIM resolution at 4.3 μm RMS. This is in good agreement with
the simulated prediction of 4-5 μm. The predicted magnification
(16x) is also in good agreement with the measured 20x.