| Neutrons are
generated at RARAF using nuclear reactions in thin targets and thus are
essentially monoenergetic, in contrast to neutrons generated by reactors or
by high energy deuterons bombarding beryllium targets. RARAF's neutron
production targets are hydrogen isotopes absorbed into thin titanium
coatings on water-cooled copper backings. Monoenergetic neutrons with
energies from 15 MeV down to 220 keV are available as shown below. Also
currently available are low-dose rate lower energy spectra. As discussed
in the section on slow neutrons, even lower-energy neutron beams (<40 keV) are available.
Learn more:
Neutron Beam
Large numbers of 14 MeV neutrons can be
produced using the T(d,n)4He reaction. The neutron energy,
fluence, and dose rate are nearly independent of angle so that planning
irradiations and designing fixtures to hold samples are relatively easy. A
significant fraction of the energy deposited in tissue by 14 MeV neutrons
is from alpha particles and heavy-ion recoils. Approximately 70% of the
energy deposition is from proton recoils.
Currently available
neutron parameters
|
Neutron energy
(MeV)
& spread (±%) |
Max. dose rate at 100
mm (Gy/hr) |
Gamma-ray
dose (%) |
Production reaction |
Incident
ion energy
(MeV) |
Angle |
|
0.11 spectrum |
0.02 |
2 |
T(p,n)3He |
1.4 |
l00° |
|
0.22 (25) |
0.20 |
1 |
T(p,n)3He |
2.0 |
120° |
|
0.3 (15) |
0.35 |
1 |
T(p,n)3He |
2.5 |
130° |
|
0.44 (14) |
0.5 |
1 |
T(p,n)3He |
2.65 |
120° |
|
0.67 (14) |
0.6 |
2 |
T(p,n)3He |
2.8 |
100° |
|
1.0 (11) |
1.1 |
1 |
T(p,n)3He |
1.9 |
0° |
|
1.5 (10) |
0.8 |
2 |
T(p,n)3He |
2.8 |
45° |
|
1.9 (4) |
2.5 |
2 |
T(p,n)3He |
2.8 |
0° |
|
2.8 (5) |
1.6 |
3 |
T(p,n)3He |
4.0 |
30° |
|
6.0 (6) |
7.0 |
6 |
D(d,n)3He |
3.1 |
0° |
|
13-15 (1-4) |
10.0 |
6 |
T(d,n)4He |
0.6
(1.8 d3) |
0°-130° |
The D(d,n)3He reaction with a
Q-value of 3.3 MeV is used to produce 6 MeV neutrons by bombarding
deuterium targets with 3.1 MeV deuterons. High dose rates are obtained by
using thicker targets. The neutron energy varies more strongly with angle
and deuteron energy than for the T(d,n) reaction because more of the
available energy is coming from the incident deuterons. The reaction cross
section has a strong forward peak so that irradiations are performed at
angles near 0° in order to maximize dose rate and reduce the fractional
dose contribution from gamma rays. At this neutron energy, about 90% of
the energy deposition is from protons, the contribution from heavy recoils
is less than for 14 MeV neutrons, and very little deposition is from alpha
particles.
Irradiations with low-energy neutrons are
performed using the T(p,n)3He reaction with a Q-value of -0.7
MeV. The neutron energy varies quite strongly with the reaction angle and
the incident particle energy. For neutron energies below 0.8 MeV,
irradiations are conducted at angles between 100° and 130° from the
incident beam direction in order to maximize dose rate and minimize energy
spread. Low-energy neutrons with energy near 440 keV are biologically most
effective because almost all of the energy is deposited by protons near
the Bragg peak which have high LETs.
A new facility for the production of useful
dose rates of neutrons with dose-mean neutron energies as low as 25 keV
(slow neutrons) has been developed using
the 7Li(p,n)7Be reaction. A water-cooled rotating
target is employed to prevent evaporation of the lithium by beam currents
of 100 µA.
Neutron Irradiation Fixtures
Neutron irradiations are performed in the
SH and SV caves. Normally, a horizontal charged-particle beam is used but
a vertical beam is available for irradiating biological systems for which
it may be better suited. Neutron irradiation fixtures for radiobiology and
physics are designed in consultation with the experimenter to meet the
needs of both the researcher and the dosimetrist.
Although the target assemblies were
designed to minimize absorbing material, the neutron dose at large angles
from the beam direction is not azimuthally uniform. To irradiate large
numbers of samples uniformly, most fixtures provide a means of rotation
about the beam axis. We have a Ferris wheel-like fixture used to irradiate
rats and mice, flasks, test tube and dishes.

Above, the
wheel has been modified for irradiation of cell monolayers growing in
commercial cell culture flasks. This arrangement has been used in
studies of transformation induction in mouse cells.
Higher dose rates to cells are delivered
using a fixture mounted on the vertical beam line. This apparatus was
originally used for irradiating hamster cells in small vials made from
plastic pipettes.
Dosimetry
Dosimetry for neutron irradiations is
performed using tissue-equivalent (TE) ionization chambers for total dose
measurements and neutron-insensitive dosimeters to measure gamma-ray dose.
The dosimetry measurements are relative to the response of a TE ionization
chamber in a fixed location which is used as a monitor. Radiation doses
are then delivered based on the response of the fixed monitor chamber. The
gamma-ray dose and the incident beam current on the target are also
monitored.
The total dose ionization chambers have
walls made of A-150 muscle TE plastic and have methane-based TE gas sealed
inside or flowing through. Insulators are made of tissue-like materials
such as styrene or Lucite. The chamber is placed at the same position as
the sample would be during the irradiation. A chamber is selected so that
the chamber volume is similar to the sample volume, and the wall thickness
is adjusted to match the amount of material between the center of the
sample and the neutron-producing target. Chambers of various sizes and
geometries from 1/4" diameter spherical to 1" diameter by 3/16" deep
parallel plate arrangements are available so that measurements can be made
even for small volumes or large areas.
If the sample is rotated about the beam
axis or if there are several samples being irradiated simultaneously,
measurements are made at various positions around the target.
Gamma-ray dosimetry is performed using
either a compensated Geiger-Mueller dosimeter or an aluminum-walled,
argon-filled (Al-Ar) ionization chamber. The large response of the Al-Ar
chamber to 14 MeV neutrons makes it unfeasible to use for that energy.
Measurements of gamma-ray dose often cannot be made at the sample position
with the G-M type dosimeter but are made as close to it as possible. The
gamma-ray dose rate is essentially isotropic about the target so that only
inverse square law corrections are necessary.
The dosimeters are calibrated using either
a 50 mg radium-226 or 7 Ci cesium-137 gamma-ray source, both of which have
been calibrated by the National Bureau of Standards.
Corrections to the dosimetry measurements
are made for any positional differences, for dose buildup or attenuation
factors due to differences in material thickness, for variations of Wn
with neutron energy, and for neutron kerma difference due to composition
differences between the samples and the chamber. Computer calculations of
the mean neutron energy, energy spread, and the relative dose rate at
various points on the sample are provided.
The electrometers used to measure the
currents (typically of the order of 10-9-10-13
A) from the ionization chambers were designed by R. Mills and
fabricated at RARAF. The higher-impedance circuitry has been potted to
minimize parasitic volume (shows almost no response to radiation) and is
attached directly to a chamber to eliminate sensitivity to movement. The
low-impedance control circuitry is located in a NIM module at the console.
|